www.rkiinstruments.com61-1006RK/61-0192RKCombustible Gas DetectorOperator’s ManualPart Number: 71-0228RKRevision: P1Released: 9/26/11
6 • 61-1006RK/61-0192RK Combustible Gas DetectorStart UpThis section describes procedures to start up the combustible gas detector and place the dete
61-1006RK/61-0192RK Combustible Gas Detector • 7MaintenanceThis section describes maintenance procedures. It includes preventive maintenance, trouble
8 • 61-1006RK/61-0192RK Combustible Gas DetectorNOTE: If the reading is not within ± 20% of the gas concentration, calibrate the detector as describe
61-1006RK/61-0192RK Combustible Gas Detector • 9Replacing the IR LEL Detector1. Turn off the controller.2. Turn off power to the controller.3. Remove
10 • 61-1006RK/61-0192RK Combustible Gas DetectorPreparing for Calibration1. Screw the calibration cup onto the detector.2. Screw the regulator into
61-1006RK/61-0192RK Combustible Gas Detector • 11Parts ListTable 3 lists replacement parts and accessories for the 61-1006RK combustible gas detector
61-1006RK/61-0192RK Combustible Gas DetectorWARNINGRead and understand this instruction manual before operating detector. Improper use of the detecto
61-1006RK/61-0192RK Combustible Gas DetectorProduct WarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in
61-1006RK/61-0192RK Combustible Gas DetectorTable of ContentsOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
61-1006RK/61-0192RK Combustible Gas Detector • 1OverviewThis manual describes the 61-1006RK combustible gas detector. This manual also describes how
2 • 61-1006RK/61-0192RK Combustible Gas DetectorDescriptionThe 61-1006RK combustible gas detector is calibrated to methane. The detector is an infrar
61-1006RK/61-0192RK Combustible Gas Detector • 3InstallationThis section describes procedures to mount the combustible gas detector in the monitoring
4 • 61-1006RK/61-0192RK Combustible Gas DetectorWiring the Combustible Gas Detector to a ControllerWARNING: Always verify that the power to the contr
61-1006RK/61-0192RK Combustible Gas Detector • 5Figure 2: Wiring the Combustible Gas Detector to a ControllerController HousingController LEL Detect
Comentários a estes Manuais