www.rkiinstruments.com Eagle 2 Maintenance Data LoaderOperator’s ManualPart Number: 71-0191RKRevision: DReleased: 5/29/14
7 • Launching the ProgramLaunching the Program 1. Click Start in the Windows® Icon Tray, then select Programs, then select Eagle 2 Maintenance Data L
Connecting an Eagle 2 to the Maintenance Data Loader Program • 8Connecting an Eagle 2 to the Maintenance Data Loader ProgramFollow these steps to con
9 • Connecting an Eagle 2 to the Maintenance Data Loader Program 3. Press and hold the POWER ENTER RESET button on the Eagle 2 until you hear a beep,
Control Buttons • 10Control ButtonsThere are 5 control buttons on the main screen: Get Current EAGLE 2 Data, Select Distributed File, Transmit New Da
11 • Control Buttons 3. Click Get Current EAGLE 2 Data to retrieve the instrument’s user and station ID, catalytic combustible user defined gases, an
Control Buttons • 12Transmit New DataAfter information has been updated in the parameter tabs, this button is used to upload that data to the Eagle 2
13 • Station & User TabStation & User TabThe Station & User tab displays a list of Station IDs and User IDs. The first time the Eagle 2 i
Station & User Tab • 14CSV FilesTo edit the Station ID or User ID list, you will need to import a csv (comma separated values) file into the prog
15 • Station & User TabThe csv files can be opened, edited, and saved using a word processing program such as Word, WordPad, or Notepad. The Stat
Station & User Tab • 16 4. Open the saved files in Word, WordPad, or Notepad, enter new Station or User IDs, and save the files. 5. In the Statio
WarrantyWarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials and workmanship, and performance f
17 • Conversion Table TabConversion Table TabThe Conversion Table tab is used to view the pre-defined relative response gases for the catalytic combu
Conversion Table Tab • 18may contain any character in upper- or lower-case.•FactorThis value is the response factor for the listed gas relative to me
19 • Conversion Table Tab• VoltThis value determines the sensor voltage for the catalytic combustible sensor. It can be set to 1.3 or 2.4. When set t
Conversion Table Tab • 20blank. Any previously defined gases will appear. 3. Export the current data by pressing the “Export csv file” button. Choose
21 • Conversion Table TabData Loader Program. The first value is the gas number, the second is the gas name, etc. 6. Edit the values you wish to chan
PID Sensor Tab • 22PID Sensor TabThe PID Sensor tab is used to view the pre-defined relative response gases and to view or define the 1 user-defined
23 • PID Sensor Tab• Full ScaleThis is the full scale value for the target gas.•PointThe point value indicates to what decimal place the gas readings
PID Sensor Tab • 24There are limitations for the full scale and increment values that depend on the factor for both the high range and the low range.
25 • PID Sensor TabWhile the pre-defined gases may not be edited, the 1 user-defined gas may be edited by doing the following: 1. With Maintenance Da
PID Sensor Tab • 26NOTE: Make sure you have downloaded the data from the EAGLE 2 to avoid erasing its data when updating the instrument using the Tra
Table of ContentsTable of ContentsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
27 • PID Sensor Tab 4. Open the csv file using Notepad, Word, or WordPad. The example below shows a csv file opened in WordPad. The user defined PID
Obtaining a Relative Response Factor • 28correct, transmit the new data to the connected instrument by pressing “Transmit New Data”. Obtaining a Rela
29 • Obtaining a Relative Response Factorcompound)• CAS No.-Sometimes it is easier to identify a VOC from the internationally recognized CAS (Chemica
Obtaining a Relative Response Factor • 30Gas/ VOC CAS No. Formula Relative Response Biphenyl 92-52-4 C12H10 0.4Bis(2,3-epoxypropyl) ether 7/5/38
31 • Obtaining a Relative Response FactorGas/ VOC CAS No. Formula Relative Response Chloro-1,3-butadiene, 2- 126-99-8 C4H5Cl 3.2Chloro-1-fluoroe
Obtaining a Relative Response Factor • 32Gas/ VOC CAS No. Formula Relative Response Dichloro-1,1-difluoroethane, 1,2-1649-08-7 C2H2Cl2F2 ZR Dichlo
33 • Obtaining a Relative Response FactorGas/ VOC CAS No. Formula Relative Response Dimethyl cyclohexane, 1,2- 583-57-3 C8H16 1.1Dimethyl disulp
Obtaining a Relative Response Factor • 34Gas/ VOC CAS No. Formula Relative Response Ethyl hexanoate 123-66-0 C8H16O2 2.6Ethyl hexanol, 2- 105-7
35 • Obtaining a Relative Response FactorGas/ VOC CAS No. Formula Relative Response Hydroquinone 123-31-9 C6H6O2 0.8Hydroxypropyl acrylate 2- 999-
Obtaining a Relative Response Factor • 36Gas/ VOC CAS No. Formula Relative Response Methoxypropyl acetate 108-65-6 C6H12O3 1.2Methyl acetate 7
1 • IntroductionIntroductionCAUTION: Read and understand this manual before using the Eagle 2 Maintenance Data Loader Program. Also read and understa
37 • Obtaining a Relative Response FactorGas/ VOC CAS No. Formula Relative Response Nitropropane, 2- 79-46-9 C3H7NO2 ZR Nitrous oxide 10024-97-
Obtaining a Relative Response Factor • 38Gas/ VOC CAS No. Formula Relative Response Propylene oxide 75-56-9 C3H6O 7Propyleneimine 75-55-8 C3H7N
39 • Obtaining a Relative Response FactorGas/ VOC CAS No. Formula Relative Response Tributylamine 102-82-9 C12H27N 1Trichloro-1,1-difluoroethane,
Spare Parts List • 40Spare Parts ListTable 5: Spare Parts ListPart Number Description47-5027RK Cable, USB/IrDA adapter71-0154RK EAGLE 2 Operator’s Ma
System Requirements • 2next section.CAUTION: The Eagle 2 detects oxygen deficiency and elevated levels of oxygen, combustible gases, carbon monoxide,
3 • Installing the Eagle 2 Maintenance Data Loader ProgramInstalling the Eagle 2 Maintenance Data Loader Program 1. Launch Windows®. 2. Exit from all
IrDA Downloading Cable • 4 5. Follow the on-screen instructions in the InstallShield Wizard Window to install the program. 6. If the InstallShield Wi
5 • IrDA Downloading CableInstalling an IrDA Adapter CableAfter installing the Eagle 2 Maintenance Data Loader Program, connect the IrDA/USB cable to
IrDA Downloading Cable • 6 6. Click OK. 7. Close the Control Panel window. Figure 2: Image Transfer Tab