Página 1 - Operator’s Manual

www.rkiinstruments.com35-3001A-11Hydrogen Sample-Draw DetectorOperator’s ManualPart Number: 71-0334Revision: P2Released: 9/17/14

Página 2

6 • 35-3001A-11 Hydrogen Sample-Draw DetectorZero PotThe zero pot is located in the upper right corner of the amplifier (see Figure 1). Use a small f

Página 3 - Product Warranty

35-3001A-11 Hydrogen Sample-Draw Detector • 7InstallationThis section describes procedures to mount the sample-draw detector in the monitoring envir

Página 4 - Table of Contents

8 • 35-3001A-11 Hydrogen Sample-Draw Detector2. Close and latch the housing door.NOTE: The sample-draw detector is shipped with the mounting feet “tu

Página 5 - Specifications

35-3001A-11 Hydrogen Sample-Draw Detector • 9ring, if it came out, in the orientation shown in Figure 4. 4. Screw the inlet fitting tube nut back on

Página 6 - Description

10 • 35-3001A-11 Hydrogen Sample-Draw DetectorInstalling the Exhaust Line1. Loosen the nut on the exhaust fitting until 3 threads are visible.2. Push

Página 7 - Internal Components

35-3001A-11 Hydrogen Sample-Draw Detector • 118. Connect the wires to the applicable detector/transmitter terminal strip at the controller as shown

Página 8

12 • 35-3001A-11 Hydrogen Sample-Draw DetectorStart UpThis section describes procedures to start up the sample-draw detector and place the sample-dra

Página 9 - Pump Reset Switch

35-3001A-11 Hydrogen Sample-Draw Detector • 136. If necessary, use a small flat-blade screwdriver to adjust the zero potentiometer until the voltmet

Página 10 - Test Points

14 • 35-3001A-11 Hydrogen Sample-Draw Detector4. Use the following formula to determine the correct test points output for the test sample.Output (mV

Página 11 - Installation

35-3001A-11 Hydrogen Sample-Draw Detector • 154. Calibrate the sample-draw detector as described in “Calibration” on page 18.5. If the fail conditio

Página 12 - Back Ferrule

35-3001A-11 Hydrogen Sample-Draw DetectorWARNINGRead and understand this instruction manual before operating detector. Improper use of the detector c

Página 13

16 • 35-3001A-11 Hydrogen Sample-Draw Detector9. Turn on power to the controller and place it into normal operation.10. Turn on the controller.CAUTIO

Página 14

35-3001A-11 Hydrogen Sample-Draw Detector • 17NOTE: When the sample-draw detector is first powered up with a new amplifier, the initial output may b

Página 15 - PATS101PATS101

18 • 35-3001A-11 Hydrogen Sample-Draw DetectorIf the low flow alarm is set too low, turn the pressure switch adjustment screw slightly clockwise. Rep

Página 16

35-3001A-11 Hydrogen Sample-Draw Detector • 19Setting the Zero Reading1. Connect the sample tubing from the demand flow regulator to the sample-draw

Página 17 - Maintenance

20 • 35-3001A-11 Hydrogen Sample-Draw DetectorParts ListTable 2 lists replacement parts and accessories for the sample-draw hydrogen detector.Table 4

Página 18 - Troubleshooting

35-3001A-11 Hydrogen Sample-Draw DetectorProduct WarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in ma

Página 19 - Sample-Draw Detector

35-3001A-11 Hydrogen Sample-Draw DetectorTable of ContentsOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 20

35-3001A-11 Hydrogen Sample-Draw Detector • 1OverviewThis operator’s manual describes the 35-3001A-11 hydrogen sample-draw detector. This manual als

Página 21

2 • 35-3001A-11 Hydrogen Sample-Draw DetectorDescriptionThis section describes the components of the hydrogen sample-draw detector. Figure 1: Hydro

Página 22 - Calibration

35-3001A-11 Hydrogen Sample-Draw Detector • 3External ComponentsThis section describes the sample-draw detector’s external components.HousingThe sam

Página 23 - Returning to Normal Operation

4 • 35-3001A-11 Hydrogen Sample-Draw DetectorMain Circuit BoardThe main circuit board includes the detector/amp terminal strip, interconnect terminal

Página 24 - Parts List

35-3001A-11 Hydrogen Sample-Draw Detector • 5counterclockwise to increase the flow and clockwise to decrease the flow.Status LEDsTwo status LEDs are