Rki-instruments M2A Manual do Utilizador Página 24

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 23
19 M2A Transmitter Operator’s Manual
Oxygen Detector, Partial Pressure, Capillary Type, and CSA Capillary Type
The wiring for all the oxygen detectors is the same. White wire to terminal labeled OXY
+, green wire to terminal labeled OXY -.
Figure 19: Partial Pressure Oxygen Detector Wiring
Figure 20: Capillary Oxygen Detector Wiring
Figure 21: CSA Capillary Type Oxygen Detector Wiring
White
Green
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
White
Green
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
White
Green
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
Vista de página 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 74 75

Comentários a estes Manuais

Sem comentários