Rki-instruments SD-705EC Manual do Utilizador Página 31

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 30
PT2E-1140
Zero adjustment impossible
Zero adjustment impossible
Yes
Yes
Zero adjustment mode
with fresh air supplied?
No
Carry out zero adjustment
while supplying the fresh air.
No
Is voltage between TP6(+) and
TPO(-) within -120mV +120mV?
Sensor deterioration.
Since the sensor must be
replaced, contact our local
agent or RIKEN KEIKI.
Yes
There may be an abnormality in
the electric circuit.
Turn OFF the power supply to
t
his unit and contact our local
agent or RIKEN KEIKI.