www.rkiinstruments.com
Eagle 2 Data Logger
Management Program
Operators Manual
Part Number: 71-0170RK
Revision: C
Released: 6/12/12
Vista de página 0
1 2 3 4 5 6 ... 119 120

Resumo do Conteúdo

Página 1 - Operator’s Manual

www.rkiinstruments.com Eagle 2 Data LoggerManagement ProgramOperator’s ManualPart Number: 71-0170RKRevision: CReleased: 6/12/12

Página 2 - Warranty

6 • IrDA Downloading CableWindows® Wireless Link Operation NoteWhen using an IrDA adapter cable and the Eagle 2 Data Logger Management Program on a W

Página 3 - Table of Contents

96 • Detail Settings Buttonfrom Table 5 on page 102 or through testing as described in “Obtaining a Relative Response Factor” on page 101. A valid ch

Página 4

Detail Settings Button • 970.50-0.59 25.00 0.010.60-0.79 30.00 0.020.80-0.99 40.00 0.021.00-1.99 50.00 0.022.00-2.99 100.0 0.13.00-3.99 150.0 0.14.00

Página 5 - Introduction

98 • Detail Settings ButtonWhile the pre-defined gases may not be edited, the 1 user-defined gas may be edited by doing the following: 1. With Data L

Página 6 - System Requirements

Detail Settings Button • 99 3. Export the current data by pressing the “Export csv file” button. Choose the file path you wish to save the file in. 4

Página 7

100 • Detail Settings Button 6. Edit the values you wish to change and save the file. 7. Return to the Data Logger Management Program with the High R

Página 8

Detail Settings Button • 101Obtaining a Relative Response FactorIf the gas that you want to monitor on the catalytic combustible or PID channel is no

Página 9 - IrDA Downloading Cable

102 • Detail Settings Buttonnumber• Formula-Molecular formula for each VOC• Response Factor (RF)-The relative response factor for each gas. This is t

Página 10 - Wireless Link Operation Note

Detail Settings Button • 103Gas/ VOC CAS No. Formula Relative Response Biphenyl 92-52-4 C12H10 0.4Bis(2,3-epoxypropyl) ether 7/5/38 C6H10O3 3Bor

Página 11 - Deselect

104 • Detail Settings ButtonGas/ VOC CAS No. Formula Relative Response Chloro-1,3-butadiene, 2- 126-99-8 C4H5Cl 3.2Chloro-1-fluoroethane, 1- 161

Página 12 - Launching the Program

Detail Settings Button • 105Gas/ VOC CAS No. Formula Relative Response Dichloro-1,1-difluoroethane, 1,2-1649-08-7 C2H2Cl2F2 ZR Dichloro-1,2,2-trif

Página 13 - Control Buttons

IrDA Downloading Cable • 7 6. Click OK. 7. Close the Control Panel window. Figure 2: Image Transfer TabDeselect

Página 14 - Instrument Information Button

106 • Detail Settings ButtonGas/ VOC CAS No. Formula Relative Response Dimethyl cyclohexane, 1,2- 583-57-3 C8H16 1.1Dimethyl disulphide 624-92-

Página 15 - Data Button

Detail Settings Button • 107Gas/ VOC CAS No. Formula Relative Response Ethyl hexanoate 123-66-0 C8H16O2 2.6Ethyl hexanol, 2- 105-76-7 C8H18O

Página 16 - 12 • Control Buttons

108 • Detail Settings ButtonGas/ VOC CAS No. Formula Relative Response Hydroquinone 123-31-9 C6H6O2 0.8Hydroxypropyl acrylate 2- 999-61-1 C6H10O3

Página 17 - Last Calibration Button

Detail Settings Button • 109Gas/ VOC CAS No. Formula Relative Response Methoxypropyl acetate 108-65-6 C6H12O3 1.2Methyl acetate 79-20-9 C3H6O

Página 18 - Set Button

110 • Detail Settings ButtonGas/ VOC CAS No. Formula Relative Response Nitropropane, 2- 79-46-9 C3H7NO2 ZR Nitrous oxide 10024-97-2 N2O ZR No

Página 19 - Downloading Data from the

Detail Settings Button • 111Gas/ VOC CAS No. Formula Relative Response Propylene oxide 75-56-9 C3H6O 7Propyleneimine 75-55-8 C3H7N 1.3Pyridine

Página 20

112 • Detail Settings ButtonGas/ VOC CAS No. Formula Relative Response Tributylamine 102-82-9 C12H27N 1Trichloro-1,1-difluoroethane, 1,2,2-354-21-

Página 21

Changing the Appearance of the Program Screens • 113Changing the Appearance of the Program ScreensTo change how information is displayed in the progr

Página 22

114 • Changing the Appearance of the Program Screens 3. Specify new fonts by clicking the button with the name of a font inscribed on it. This action

Página 23

Changing the Appearance of the Program Screens • 115 4. Select the colors used to graph the various target gases in the interval trend and alarm tren

Página 24 - Eagle 2 Data Logging Capacity

8 • Launching the ProgramLaunching the Program 1. Click Start on the Windows® Icon Tray, then select All Programs, then select Eagle 2. Your operatin

Página 25

116 • Spare Parts ListSpare Parts ListTable 6: Spare Parts ListPart Number Description47-5027RK Cable, USB/IrDA adapter71-0170RK Eagle 2 Data Logger

Página 26 - Information Screen

Control Buttons • 9Control ButtonsThis section provides an overview of the control buttons. Instructions for using the various parts of the program a

Página 27

10 • Control ButtonsDownload ButtonIf you are in another program window, clicking the Download button opens the Download Window (see Figure 3). The D

Página 28 - Data Window

Control Buttons • 11commands and is currently connected to the program. If an instrument is turned off after being connected to the program, then the

Página 29

12 • Control Buttons• Interval Trend Data FilesInterval trend data is logged at the interval time defined in the Eagle 2. Each logged point is an ave

Página 30 - Calibration History

Control Buttons • 13Last Calibration ButtonClicking the Last Calibration button opens the Last Calibration Window.The Last Calibration Window stores

Página 31

14 • Control ButtonsSet ButtonClicking the Set button opens the Set Window.In the Set Window, you can perform the following functions:• Alter the app

Página 32

Downloading Data from the Eagle 2 • 15Downloading Data from the Eagle 2You have the option of downloading data manually or automatically. If you want

Página 33

WarrantyWarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials and workmanship, and performance f

Página 34 - Event Data

16 • Downloading Data from the Eagle 2IrDA adapter cable as shown in Figure 10 below, aligning the infrared port on the front of the Eagle 2 with the

Página 35

Downloading Data from the Eagle 2 • 17Power Off Download Commands become selectable. 5. If you are going to download data manually, you can perform a

Página 36

18 • Downloading Data from the Eagle 2 7. If the unit has been downloaded manually (Automatic Download not selected), the unit’s instrument informati

Página 37

Downloading Data from the Eagle 2 • 19 8. You can view, print, export, or delete downloaded data by entering the Data or Last Calibration Windows. T

Página 38 - Interval Trend Data

20 • Eagle 2 Data Logging CapacityWARNING: If you click Clear Logger Data, all data is erased in the Eagle 2, but not in your computer’s memory. So i

Página 39

Overwriting Data in the Eagle 2 • 21Overwriting Data in the Eagle 2The Eagle 2’s Data Log Overwrite function is factory set to On so that when the Ea

Página 40

22 • Viewing Data in the Instrument Information ScreenViewing Data in the Instrument Information ScreenYou can view information for an instrument tha

Página 41

Viewing Data in the Instrument Information Screen • 23• Gas - Lists each channel’s gas in the order it appears on the Eagle 2 screen.• Calib. Date -

Página 42

24 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowViewing, Printing, Exporting, and Deleting Data in the Data WindowThe Eagle 2

Página 43

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 25The Data Window is divided into four frames. The upper left frame is the Data

Página 44

Table of ContentsTable of ContentsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Sys

Página 45

26 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowCalibration HistoryThe Eagle 2 is capable of saving calibration information fo

Página 46

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 27To view, print, or export the calibration history for any instrument in the da

Página 47

28 • Viewing, Printing, Exporting, and Deleting Data in the Data Windowbelow the untitled folder along with the Alarm Events and Trouble Events folde

Página 48 - Cursor FeatureEvent Select

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 29 9. If you want to print the data, click the Print button. A Printer List dial

Página 49

30 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowAfter specifying the file name, file type, and file location click the Save bu

Página 50

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 31To view, print, or export the event data for any instrument in the database: 1

Página 51

32 • Viewing, Printing, Exporting, and Deleting Data in the Data Windowuntitled folder. The Alarm Events Folder and Trouble Events Folder will appear

Página 52

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 33 8. If you click the Summary button, the Data Window will split into two frame

Página 53

34 • Viewing, Printing, Exporting, and Deleting Data in the Data Window 11.To export the data for use in another application, for example a spreadshe

Página 54 - Alarm Trend Data

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 35An interval trend data file is created and saved in the Eagle 2 when the instr

Página 55

Table of ContentsChanging the Password . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63Viewing, Printing, & Deleting Data i

Página 56

36 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowTo view and perform desired operations with the interval trend files: 1. With

Página 57

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 37dated folder whose contents you want to see. 5. Click on the Interval Trend Fo

Página 58

38 • Viewing, Printing, Exporting, and Deleting Data in the Data Window• In table format, the interval trend data is shown as the average gas reading

Página 59 - Cursor feature

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 39• The catalytic combustible channel can be recorded in %LEL, ppm, or %volume u

Página 60

40 • Viewing, Printing, Exporting, and Deleting Data in the Data Window• If installed, an infrared methane or hydrocarbon channel records data in %LE

Página 61

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 41 8. When the data is viewed in table format, if you move the cursor over an al

Página 62 - Cursor feature IR HC or CH4

42 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowIf you click on the event, the corresponding alarm trend data file will be ope

Página 63

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 43• If there are more than two consecutive data points with the same readings fo

Página 64

44 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowWhen viewing interval trend data in graph format, you have several options: •

Página 65

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 45session and on the zoom factor. Depending on the length of the data session, d

Página 66

Introduction • 1IntroductionUsing an advanced detection system consisting of up to six gas sensors, the Eagle 2 Gas Monitor detects the presence of c

Página 67 - Changing the Password

46 • Viewing, Printing, Exporting, and Deleting Data in the Data Windowchannel is the first channel and is displayed in green.• If installed, an IR H

Página 68

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 47%LEL up to 100% LEL and then it will start recording data in % volume up to 10

Página 69

48 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowNOTE: The IR Range Select selection box in the lower right hand corner only ap

Página 70 - Calibration Window

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 49information shown in the lower right frame in Figure 26. To return to the one

Página 71

50 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowFor Graph view, the export file type is Windows bitmap (.bmp). For Table view,

Página 72

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 51To view and perform desired operations with the alarm trend files: 1. With the

Página 73

52 • Viewing, Printing, Exporting, and Deleting Data in the Data Window 5. Click on the Alarm Trend Folder in the upper left frame. In the upper righ

Página 74

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 53• In table format, the log times are shown along with the peak (minimum for ox

Página 75

54 • Viewing, Printing, Exporting, and Deleting Data in the Data Window• If installed, an infrared methane or hydrocarbon channel records data in %LE

Página 76

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 55 8. The Event Only and Condensed selection boxes are not selectable for alarm

Página 77 - The Eagle 2 Data Logger

2 • System RequirementsCAUTION: The Eagle 2 detects oxygen deficiency and elevated levels of oxygen, combustible gases, carbon monoxide, and hydrogen

Página 78

56 • Viewing, Printing, Exporting, and Deleting Data in the Data Windowrecorded in any of these units and stored in the same file, those readings may

Página 79

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 57range of the sensor. If the channel is configured as a %LEL only channel, then

Página 80

58 • Viewing, Printing, Exporting, and Deleting Data in the Data WindowNOTE: The IR Range Select selection box in the lower right hand corner only ap

Página 81

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 59• When selected, the Event feature shows on the graph where each channel went

Página 82

60 • Viewing, Printing, Exporting, and Deleting Data in the Data Window 12.To export the data for use in another application, for example a spreadshe

Página 83

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 61• Interval Trend and Alarm Trend data filesTo delete any of the above items in

Página 84 - Detail Settings Button

62 • Viewing, Printing, Exporting, and Deleting Data in the Data Windowthe program is first installed is “rki”. The password is case sensitive. See “

Página 85 - Detail Settings Button • 81

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 63Changing the PasswordCAUTION: Changing the password requires use of the Delete

Página 86 - Station & User Tab

64 • Viewing, Printing, Exporting, and Deleting Data in the Data Window 3. Click Change Password. The Password Window asks you to input the current p

Página 87 - Detail Settings Button • 83

Viewing, Printing, Exporting, and Deleting Data in the Data Window • 65 7. Click OK when the program confirms that you have changed the password. 8.

Página 88 - 84 • Detail Settings Button

Installing the Eagle 2 Data Logger Management Program • 3Installing the Eagle 2 Data Logger Management Program 1. Launch Windows®. 2. Exit from all a

Página 89 - Detail Settings Button • 85

66 • Viewing, Printing, and Deleting Data in the Last Calibration WindowViewing, Printing, and Deleting Data in the Last Calibration WindowYou can ac

Página 90 - Click to import User csv file

Viewing, Printing, and Deleting Data in the Last Calibration Window • 67The Need Calibration view option shows the following fields:• No. — lists, in

Página 91 - Detail Settings Button • 87

68 • Viewing, Printing, and Deleting Data in the Last Calibration Window• Last Downloaded — shows when the last download took place for a unit using

Página 92 - Conversion Table Tab

Viewing, Printing, and Deleting Data in the Last Calibration Window • 69To print a list of the instruments shown in the Calibration Date view option

Página 93 - Detail Settings Button • 89

70 • Viewing, Printing, and Deleting Data in the Last Calibration Window• Before — shows the settings prior to calibration.• After — shows the settin

Página 94 - 90 • Detail Settings Button

Viewing, Printing, and Deleting Data in the Last Calibration Window • 71 4. Click the row with the right mouse button. The Delete box appears. 5. Cli

Página 95 - Detail Settings Button • 91

72 • Viewing, Printing, and Deleting Data in the Last Calibration Window 7. When the password has been entered and the Continue button clicked, the D

Página 96 - 92 • Detail Settings Button

Calibrating an Eagle 2 With The Eagle 2 Data Logger Management Program • 73Calibrating an Eagle 2 With The Eagle 2 Data Logger Management ProgramAn E

Página 97 - Detail Settings Button • 93

74 • Calibrating an Eagle 2 With The Eagle 2 Data Logger Management Pro-CAUTION: Calibration using the Eagle 2 Data Logger Management Program should

Página 98 - PID Sensor Tab

Calibrating an Eagle 2 With The Eagle 2 Data Logger Management Program • 75 6. Click the Set button to display the Set window. 7. Click the Calibrati

Página 99 - Detail Settings Button • 95

4 • Installing the Eagle 2 Data Logger Management Program 5. Follow the on-screen instructions in the InstallShield Wizard Window to install the prog

Página 100 - 96 • Detail Settings Button

76 • Calibrating an Eagle 2 With The Eagle 2 Data Logger Management Pro- 8. Use the selection boxes under the Select column to select the channels yo

Página 101 - Detail Settings Button • 97

Changing Eagle 2 Instrument Parameters • 77 13.Connect the regulator to the Eagle 2 probe tube using the sample tubing provided with the calibration

Página 102 - 98 • Detail Settings Button

78 • Changing Eagle 2 Instrument ParametersOff download commands will be selectable. Click the Instrument Information download command to retrieve th

Página 103 - Detail Settings Button • 99

Changing Eagle 2 Instrument Parameters • 79the standard four gas section. 6. To change the serial number, click the serial number field and use the b

Página 104 - 100 • Detail Settings Button

80 • Detail Settings Button2 by using the POWER ENTER RESET button on the Eagle 2 or by clicking the Power Off button and confirming that you want to

Página 105 - Detail Settings Button • 101

Detail Settings Button • 81of Instrument Information. 5. Click the Set button to display the Set window. Figure 66: Set Window

Página 106 - 102 • Detail Settings Button

82 • Detail Settings ButtonStation & User TabThe Station & User tab displays a list of Station IDs and User IDs. The first time the Eagle 2 i

Página 107 - Detail Settings Button • 103

Detail Settings Button • 83CSV FilesTo edit the Station ID or User ID list, you will need to import a csv (comma separated values) file into the prog

Página 108 - 104 • Detail Settings Button

84 • Detail Settings Button 2. Navigate to the location you would like to save the csv file, type in a file name, and click Save. Figure 69: Station

Página 109 - Detail Settings Button • 105

Detail Settings Button • 85 3. The csv files can be opened, edited, and saved using a word processing program such as Word, WordPad, or Notepad. The

Página 110 - 106 • Detail Settings Button

IrDA Downloading Cable • 5IrDA Downloading CableThe Eagle 2 communicates with a computer via an on-board infrared communication port that complies wi

Página 111 - Detail Settings Button • 107

86 • Detail Settings Button 2. In the Station and User tab, click “Import csv file” for either the Station ID or User ID and select a csv file. Figur

Página 112 - 108 • Detail Settings Button

Detail Settings Button • 87 3. Select the file you want to import and click Open. 4. Once the Station ID and/or User ID lists have been imported, cli

Página 113 - Detail Settings Button • 109

88 • Detail Settings Button 7. Click Yes in the confirmation window that appears.Conversion Table TabThe Conversion Table tab is used to view the pre

Página 114 - 110 • Detail Settings Button

Detail Settings Button • 89There are 8 columns in both the Pre-Defined Table tab and the User-Defined Table tab:•NoThis column represents the gas num

Página 115 - Detail Settings Button • 111

90 • Detail Settings ButtonNOTE: If you define a gas whose LEL is above 50,000 ppm, the %LEL reading in Measuring Mode will reflect the defined ppm r

Página 116 - 112 • Detail Settings Button

Detail Settings Button • 91While the pre-defined gases may not be edited, the 5 user-defined gases may be edited by doing the following: 1. With the

Página 117 - Changing the Appearance

92 • Detail Settings Button 4. Open the csv file using Notepad, Word, or WordPad. The example below shows a csv file opened in WordPad. The list of g

Página 118 - Figure 85: Font Window

Detail Settings Button • 93 7. Return to the Data Logger Management Program with the User-Defined Table tab still up and press “Import csv file”. 8.

Página 119

94 • Detail Settings ButtonPID Sensor TabThe PID Sensor tab is used to view the pre-defined relative response gases and to view or define the 1 user-

Página 120 - Spare Parts List

Detail Settings Button • 95• Full ScaleThis is the full scale value for the target gas.•PointThe point value indicates to what decimal place the gas