Maintainin
g
the GX-2001 Calibration Station 56
8. Use the fresh air adjustment valve to set the ow to 1.0
SCFH.
NOTE: If the calibration station switches over to drawing span
gas before you are able to complete the adjustment, use
the span gas adjustment valve to make ow
adjustments.
9. Click the Cancel button on the screen to abort the calibration
sequence.
10.Turn off the calibration station and exit the downloading
software.