126 Using Setup Mode EAGLE 2 Operator’s Manual
4. Press and release POWER ENTER RESET to save the setting and return
to the main menu.
Zero Follower Settings
The ZERO FOLLOWER setting is not intended for field adjustment. The
default setting for most target gases is ON. The default setting for carbon
dioxide channels and some configurations of non-standard toxic gas channels
is OFF. The oxygen channel does not support this feature.
Zero Suppression Settings
The ZERO SUPRESSION setting is not intended for field adjustment. The
typical setting is 2% of full scale. The oxygen channel does not support this
feature.
Updating the Confirmation Alert Setting
With CONFIRMATION ALERT set to BEEP AND LIGHT, the EAGLE 2
beeps and flashes the LED arrays once every 15 minutes to verify that it is
operating.
With CONFIRMATION ALERT set to LIGHT ONLY, the EAGLE 2
flashes the LED arrays once every 15 minutes to verify that it is operating.
With CONFIRMATION ALERT set to BEEP ONLY, the EAGLE 2 beeps
once every 15 minutes to verify that it is operating.
With CONFIRMATION ALERT set to OFF (factory setting), the EAGLE
2 does not sound a confirmation alert.
1. From the main menu, place the cursor next to CONFIRMATION
ALERT.
2. Press and release POWER ENTER RESET. The Confirmation Alert
Screen appears.
3. Use AIR YES or RANGE SHIFT to display the desired setting.
4. Press and release POWER ENTER RESET to save the setting and return
to the main menu.
CONFIRMATION ALERT
UP/DOWN THEN ENTER
OFF