Rki-instruments EAGLE 2 Manual Manual do Utilizador Página 282

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 281
272 Start Up, Bar Hole Mode EAGLE 2 Operator’s Manual
NOTE: If no button is pressed for 20 seconds, the unit will proceed
into whichever mode has the cursor in front of it.
9. The Active Gases Screen appears for a few seconds indicating which
channels are active and their target gas.
10. The gas alarm setpoints are displayed by three screens in sequence: the
Low Alarm Screen, High Alarm Screen, and STEL/TWA Alarm Screen.
Each screen remains on the LCD for three seconds.
11. After the alarm screens, if CAL REMINDER is turned on, the screen that
appears next depends on how CAL PAST DUE ACT is set in the Setup
Mode Menu (see “Updating the Calibration Past Due Action Setting” on
page 123).
ACTIVE GASES
CH4 OXY
H2S CO
CH4
A CH4 10 %LEL
L L OXY 19.5 vol%
O A H2S 10.0 ppm
WR CO 25 ppm
M
S
CH4 10 vol%
A CH4 50 %LEL
HL OXY 23.5 vol%
I A H2S 30.0 ppm
GR CO 50 ppm
HM
S
CH4 50 vol%
ALARMS STEL & TWA
H2S(ppm) 15.0 10.0
CO (ppm) 200 25
CH4(vol%) OFF OFF
Vista de página 281
1 2 ... 277 278 279 280 281 282 283 284 285 286 287 ... 308 309

Comentários a estes Manuais

Sem comentários