GX-2012 Operator’s Manual Display Mode • 45
The lunch break feature enables the GX-2012 to remember peak readings
when it is turned off so it can continue them when it is turned on again. See
“Turning On the GX-2012, Normal Mode Only” on page 19. To reset the
peak readings during operation, press and hold the RESET SILENCE button
for a few seconds while in the peak screen.
STEL Screen
The STEL Screen displays the short-term exposure limit (STEL) readings for
CO and H
2
S only. The STEL reading is the average reading during the last 15
minutes.
TWA Screen
The TWA Screen displays the time-weighted average (TWA) readings for CO
and H
2
S only.
The TWA reading is the average reading during the last 8 hours. If 8 hours
have not elapsed since the last time the TWA reading was cleared, the
average is still calculated over 8 hours. The missing time is assigned a 0
value for readings.
0
20.9
0
0.0
%LEL
%
ppm
ppm
CH4
O2
PEAK
CO
H2S
PEAK
0
0.0
ppm
ppm
STEL
CO
H2S
0
0.0
ppm
ppm
TWA
CO
H2S
Vista de página 52
1 2 ... 48 49 50 51 52 53 54 55 56 57 58 ... 184 185

Comentários a estes Manuais

Sem comentários

Intellinet Adaptery PoE podręczniki użytkownika

Lista Intellinet Adaptery PoE produkty zawierają 4 instrukcje obsługi do modeli 4. Wszystkie podręczniki użytkownika w formacie pdf są dostępne do pobrania lub obejrzenia online.

Kliknij na poniższy alfabet, aby przejść do pełnej listy modeli zaczynających się od tej litery

Modele Rodzaj dokumentu
502900
Instrukcja Użytkownika       Intellinet 502900, 8 strony
502931
Instrukcja Użytkownika        Intellinet 502931, 12 strony
560436
Instrukcja Użytkownika      Intellinet 560436, 8 strony
PoE Injector
Instrukcja Użytkownika       Intellinet PoE Injector, 8 strony