www.rkiinstruments.com SDM-E2 Docking StationPC Controlled ConfigurationOperator’s ManualPart Number: 71-0251RKRevision: EReleased: 5/29/14

Página 2 - Warranty

5 • DescriptionDescriptionThis section describes the SDM-E2 docking station. It is designed to be used on a table top and consists of the AC adaptor,

Página 3 - Table of Contents

95 • Calibrating an Instrument with Special Sensors 12. After the GAS 1 calibration gas is applied, calibration gas for the first special sensor will

Página 4

Calibrating an Instrument with Special Sensors • 96 13. If you are calibrating an instrument or instruments with 2 special sensors, you will then be

Página 5

97 • Calibrating an Instrument with Special Sensors Figure 73: Second GAS 2 FlowingThe PC Controller Program will analyze the results.

Página 6 - Introduction

Calibrating an Instrument with Special Sensors • 98 14. After calibration gas is applied, the docking stations will purge with fresh air for 45 secon

Página 7 - System Requirements

99 • Calibrating an Instrument with Special Sensors 15. After the fresh air purge is complete,• If the calibration passes, the following window will

Página 8 - Specifications

Calibrating an Instrument with Special Sensors • 100• If the calibration fails, the following window will open indicating which channels failed to ca

Página 9 - About this Manual

101 • Calibrating an Instrument with Special Sensors• If at any point during the calibration the flow rate to the instrument becomes too low, the PC

Página 10 - Description

Troubleshooting • 102TroubleshootingNOTE: This troubleshooting guide describes SDM-E2 problems only. See the EAGLE 2 Operator’s Manual for problems y

Página 11 - USB Cable

103 • Charging an Instrument in an SDM-E2Charging an Instrument in an SDM-E2The SDM-E2 can be used to charge the rechargeable batteries in an EAGLE 2

Página 12 - Instrument Panel

Charging an Instrument in an SDM-E2 • 104 2. After the bump test or calibration has been completed, turn off the instrument(s) by selecting the instr

Página 13 - Back Panel

Description • 6USB CableA Type A to Type B USB cable is provided with the docking station. It is used to connect the USB port on the back of the SDM-

Página 14 - Control Panel

105 • Charging an Instrument in an SDM-E2release the POWER button. The COPY LED will remain on if the SDM-E2 was used in Standalone Configuration and

Página 15 - Station Systems

Logs Button • 106Logs ButtonThe Logs Window contains saved data from EAGLE 2 instruments in addition to records of bump tests and calibrations perfor

Página 16 - Hardware Setup

107 • Logs ButtonTo view data in the Date Base View Format:a. Select Date in the Base view box.b. Click the expanded view symbol (+) next to a date f

Página 17

Logs Button • 108next to the item or double click it. 2. ID View FormatThe data may be organized by one or more of the following parameters: • Serial

Página 18

109 • Logs ButtonIf any of these boxes are selected, the Base view box disappears. Click the selection box or boxes in the lower left frame to organi

Página 19 - Hardware Setup • 14

Logs Button • 110same year/month, they will be distinguished by their different serial numbers.NOTE: In the examples that follow, the data will be sh

Página 20

111 • Logs Button 3. Click the year/month folder you wish to view. All alarm trend data files for that instrument in that particular year/month will

Página 21 - Hardware Setup • 16

Logs Button • 112 4. Click a file to see the instrument information, gas readings at the time of the event, and the alarm setpoints in the lower righ

Página 22

113 • Logs Button 5. To view the alarm trend data, double click the desired file. Figure 83: Alarm Trend DataEach data point will be displayed in the

Página 23

Logs Button • 114%volume units depending on the instrument setting. However, when viewing alarm trend data, the data for the catalytic LEL channel ca

Página 24 - Connecting Calibration Gas

7 • DescriptionA check valve is included with the SDM-E2. It is intended for use on the exhaust fitting when manifolding multiple docking stations to

Página 25

115 • Logs Button 7. In the graph part of the screen,• You can choose which gas(es) you want to graph by selecting or deselecting the boxes next to e

Página 26

Logs Button • 116The IR Range Select box is located in the upper right corner of the graph portion of the alarm trend screen just to the right of the

Página 27 - PC Controller Software

117 • Logs Buttonhighlight that data point in the table. Clicking a data point in the table will move the red line in the graph to that data point. 8

Página 28

Logs Button • 118Calibration DataThe EAGLE 2 is capable of saving calibration information for up to the 100 most recent calibrations. When the maximu

Página 29

119 • Logs Button 4. Click one of the calibration data files. The fields in the bottom right frame will fill in. The instrument information will be d

Página 30

Logs Button • 120Calibration Report control button along the top of the Logs window. A dialog box will appear confirming if you want to print. Click

Página 31

121 • Logs Buttonsaved in the program.Alarm event files save the time, instrument channel, gas, and alarm type of every gas alarm event that occurs o

Página 32 - PC Controller Program

Logs Button • 122selected and click the Print control button in the upper left corner of the Logs window. A dialog box will appear confirming if you

Página 33 - Controller Program

123 • Logs Buttonoperation. The serial number, user ID, station ID, and catalytic combustible channel configuration that are entered in the instrumen

Página 34 - Main Program Window

Logs Button • 124 3. Click the year/month folder you wish to view. All interval trend data files for that instrument in that particular year/month wi

Página 35 - Setting Up the Configuration

Description • 8Back PanelThe back panel includes the power jack, sample fittings, gas out to EAGLE 2 fitting, and a USB PC connector. Power JackThe p

Página 36

125 • Logs Button 4. Click one of the interval trend data files. The fields in the bottom right frame will fill in. The instrument information will b

Página 37

Logs Button • 126 5. To view the interval trend data, double click the desired file. Figure 94: Interval Trend DataIn table format on the left half o

Página 38

127 • Logs ButtonTo change the proportions of the table to the graph, click the divider between the two and move it.To display instrument information

Página 39

Logs Button • 128• If installed, an infrared methane or hydrocarbon channel records data in %LEL and/or % volume units depending on whether the chann

Página 40

129 • Logs Button%LEL is chosen as the unit, any ppm or %volume data is converted and graphed as %LEL. The user may change back and forth between the

Página 41

Logs Button • 130The IR Range Select box is located in the upper right corner of the graph portion of the interval trend screen just to the right of

Página 42 - Cylinders Window

131 • Logs Buttonpoint in the table will move the red line in the graph to that data point. 8. To print the data, select either List or Graph in the

Página 43

Logs Button • 132Memo DataAny information entered into the Memo field in the Edit window is saved in a file in the Logs window. If there is text in t

Página 44

133 • Logs Button 4. Click on one of the memo files. The fields in the bottom right screen will fill in. The instrument information will be displayed

Página 45

Logs Button • 134 6. To save memo data, click the Save control button in the upper left corner of the Logs window. A “Save As” dialog box will appear

Página 46 - 3 SEC OFF

9 • DescriptionControl PanelThe control panel is used to setup and operate the docking station in the Standalone configuration. It is located at the

Página 47

135 • Logs Button 3. Click the year/month folder you wish to view. All bump test data files for that instrument in that particular year/month will ap

Página 48

Logs Button • 136 4. Click on one of the bump test data files. The fields in the bottom right screen will fill in. The instrument information will be

Página 49 - Icon View vs. Details View

137 • Logs ButtonReport control button along the top of the Logs window. A dialog box will appear confirming if you want to print. Click OK. 10. To p

Página 50

Logs Button • 138Deleting Data in the Logs WindowData can be deleted from the Logs window by right clicking on the folder or file you wish to delete.

Página 51

139 • Instrument Function MenuInstrument Function MenuWhen you are in the main program window, if you right click on an instrument or a docking stati

Página 52 - Testing or Calibration

Instrument Function Menu • 140Sensor Tab, the Station & User Tab, the Conversion Table Tab, the PID Sensor Tab, the Channel Gas Tab, and the Gas

Página 53

141 • Instrument Function MenuSeveral parameters are always selectable and do not depend on which sub tab is selected. • Serial No.: To change the cu

Página 54

Instrument Function Menu • 142will still appear in the Memo field. 2. If necessary, delete the old memo. Enter the information you wish to document.

Página 55

143 • Instrument Function MenuOption 1 Sub TabThe adjustable parameters in the Option 1 Tab are:• OXY Alarm Pattern: The oxygen alarms can be set to

Página 56

Instrument Function Menu • 144choose to continue accumulating TWA and PEAK readings and the time in operation from the last time the EAGLE 2 was used

Página 57

USB Hub Requirements for Multiple Station Systems • 10Front PanelA type A USB port is located on the front of the docking station. This port can be u

Página 58

145 • Instrument Function MenuOption 2 Sub TabThe Option 2 Tab of the Parameter Tab contains additional parameters that can be changed.• Date Time Di

Página 59

Instrument Function Menu • 146• BH Measuring Time: The BH Measuring Time parameter only applies to instruments being used in Bar Hole Mode. The BH Me

Página 60

147 • Instrument Function MenuA confirmation screen will appear asking if you would like to save the changes you have made. To save the changes and r

Página 61

Instrument Function Menu • 148 2. Click the box again and only the value will be highlighted in blue. 3. Type in the new desired value. 4. Repeat ste

Página 62

149 • Instrument Function Menu 2. Navigate to the location you want to save the file, type in a file name, and click Save. You can save multiple stat

Página 63

Instrument Function Menu • 150 3. The csv files can be opened, edited, and saved using a word processing program such as Word, WordPad, or Notepad. T

Página 64

151 • Instrument Function MenuLoading ID Files 1. In the Station & User Tab, click Import csv file for either the Station ID list or User ID list

Página 65

Instrument Function Menu • 152 2. Navigate to the file you wish to import and click Open. 3. Once you have imported the Station ID and/or User ID csv

Página 66

153 • Instrument Function Menu 4. Click OK to save the changes and click Yes in the confirmation window that comes up. The program will tell you to w

Página 67

Instrument Function Menu • 154Conversion Table TabThe Conversion Table tab is used to view the pre-defined relative response gases for the catalytic

Página 68

11 • Hardware SetupHardware SetupThis section describes how to assemble the hardware that came with the SDM-E2, assemble a manifold, and connect cali

Página 69

155 • Instrument Function Menufactor for each pre-defined gas is factory defined. The factor for any user-defined gases must be obtained through test

Página 70

Instrument Function Menu • 156While the pre-defined gases may not be edited, the 5 user-defined gases may be edited by doing the following: 1. Click

Página 71

157 • Instrument Function Menu 4. Open the csv file using Notepad, Word, or WordPad. The example below shows a csv file opened in WordPad. The list o

Página 72

Instrument Function Menu • 158 9. The values you entered in the csv file will appear in the User-Defined Table. 10. If the program finds anything wro

Página 73

159 • Instrument Function Menu•NameThis is what will appear in the PID Sensor Relative Response Screen in the EAGLE 2. The name can be up to 3 charac

Página 74

Instrument Function Menu • 160There are limitations for the full scale and increment values that depend on the factor for both the high range and the

Página 75

161 • Instrument Function MenuWhile the pre-defined gases may not be edited, the 1 user-defined gas may be edited by doing the following: 1. Click on

Página 76

Instrument Function Menu • 162 2. The user-defined gas is gas 17. If no user-defined PID gas has been previously defined for the connected instrument

Página 77

163 • Instrument Function Menu 4. Open the csv file using Notepad, Word, or WordPad. The example below shows a csv file opened in WordPad. The user d

Página 78

Instrument Function Menu • 164Channel Gas TabThe Channel Gas Tab allows the view the gas channel order and various parameters related to those channe

Página 79

Hardware Setup • 12NOTE: If you have an EAGLE 2 with an IR CO2 sensor in the 0-10,000 ppm or 0-5.00 %vol range, be sure to replace the air filter at

Página 80

165 • Instrument Function MenuGas Name TabThe Gas Name Tab displays current gas names and values related to the catalytic LEL channel. The informatio

Página 81

Obtaining a Relative Response Factor • 166Power Off FunctionThis function can be activated by either right clicking an instrument and selecting Power

Página 82

167 • Obtaining a Relative Response FactorSome abbreviations that appear in the table are:• ZR-No response• NV-Cannot be measuredTable 7: Response Fa

Página 83

Obtaining a Relative Response Factor • 168Gas/ VOC CAS No. Formula Relative Response Butane, n- 106-97-8 C4H10 46Butanol, 1- 71-36-3 C4H10O 4B

Página 84

169 • Obtaining a Relative Response FactorGas/ VOC CAS No. Formula Relative Response Chlorotoluene, o- 95-49-8 C7H7Cl 0.5Chlorotoluene, p- 108-4

Página 85

Obtaining a Relative Response Factor • 170Gas/ VOC CAS No. Formula Relative Response Dichloroethane, 1,1- 75-34-3 C2H4Cl2 ZR Dichloroethene, 1,1

Página 86

171 • Obtaining a Relative Response FactorGas/ VOC CAS No. Formula Relative Response Dinitrobenzene, o- 528-29-0 C6H4N2O4 ZR Dinitrobenzene, p-

Página 87

Obtaining a Relative Response Factor • 172Gas/ VOC CAS No. Formula Relative Response Formic acid 64-18-6 CH2O2 ZR Furfural 98-01-1 C5H4O2 1.4Fu

Página 88

173 • Obtaining a Relative Response FactorGas/ VOC CAS No. Formula Relative Response Isodecanol 25339-17-7 C10H22O 0.9Isoflurane 26675-46-7 C3H2Cl

Página 89

Obtaining a Relative Response Factor • 174Gas/ VOC CAS No. Formula Relative Response Methyl t-butyl ether 1634-04-4 C5H12O 0.8Methyl-2-propen-1

Página 90

13 • Hardware Setupconfigurations, see “Instrument Placement for Bump Testing or Calibration” on page 47 for more information and suggestions for man

Página 91

175 • Obtaining a Relative Response FactorGas/ VOC CAS No. Formula Relative Response Pentachloroethane 76-01-7 C2HCl5 ZR Pentachlorofluoroethane 3

Página 92

Obtaining a Relative Response Factor • 176Gas/ VOC CAS No. Formula Relative Response Tetrabromoethane, 1,1,2,2- 79-27-6 C2H2Br4 2Tetracarbonylni

Página 93

177 • Config WindowConfig WindowTo enter the Configuration Window, click the Config button. The program will prompt you for a password. The default p

Página 94

Config Window • 178Database TabThe Database Tab allows you to view or change where the data from your SDM-E2 docking stations is saved. It also allow

Página 95

179 • Config Window 2. Choose where you would like the new database to be stored and enter a name for it and click Save. The new database will be cre

Página 96

Config Window • 180Password TabThe password Tab allows you to change the Configuration Window password to something other than the factory set ABCDE.

Página 97

181 • Spare Parts List 2. To cancel any changes you have made at any time, click Cancel in the main Password Tab. The program will tell you that chan

Página 98

Spare Parts List • 18281-0013RK-05 Calibration cylinder, 50% vol CH4 in N2, 58 liter steel81-0018RK-01 Calibration cylinder, 50 %LEL isobutane in air

Página 99

183 • Spare Parts List81-0142RK-02 Five-gas calibration cylinder, 50% LEL CH4/12% O2/25 ppm H2S/50 ppm CO/5 ppm SO2 in nitrogen, 58 liter aluminum81-

Página 100 -

Spare Parts List • 184814-101-P25 Calibration kit, 5-gas with IBL, 58 liter aluminum cylinder814-101-P25-DLV Calibration kit, 5-gas with IBL, 34 lite

Página 101

Hardware Setup • 14Unit 1T-FittingUnit 2Unit 9Unit 109 inches30 Foot Max Exhaust TubeTo Open AreaCheck Valve...7 inches Figure 11: Exhaust Tubing Con

Página 102

185 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura-Appendix A: Multiple Instrument ConfigurationsEAGLE 2s with different sensor com

Página 103

Bump Testing Multiple EAGLE 2s with Different Sensor Configurations • 186NOTE: If you have an EAGLE 2 with an IR CO2 sensor in the 0-10,000 ppm or 0-

Página 104

187 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura- 8. You will be asked to confirm which EAGLE 2 channels will use GAS 1 for the b

Página 105

Bump Testing Multiple EAGLE 2s with Different Sensor Configurations • 188NOTE: Do not select every gas in the gas cylinder confirmation screen unless

Página 106

189 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura- 9. You will then be asked to confirm the gas cylinder concentration and the ins

Página 107 - Troubleshooting

Bump Testing Multiple EAGLE 2s with Different Sensor Configurations • 190If you are bump testing single channel special sensor instruments and do not

Página 108

191 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura-If you are bump testing single channel special sensor instruments and have a man

Página 109

Bump Testing Multiple EAGLE 2s with Different Sensor Configurations • 192 10. Click OK. You will be asked if you want to bump test the instruments. C

Página 110 - PLEASE WAIT

193 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura- 12. After air is applied, and if all channels passed the zero adjustment, calib

Página 111 - Logs Button

Bump Testing Multiple EAGLE 2s with Different Sensor Configurations • 194 13. After GAS 1 calibration gas is applied, calibration gas for the first s

Página 112

WarrantyWarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials and workmanship, and performance f

Página 113

15 • Hardware SetupGAS 1 TubingThe GAS 1 fittings for up to 10 docking stations can be manifolded together. Reference Figure 12 for the instructions

Página 114

195 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura- 14. You will then be prompted to change the GAS 2 cylinder. The pumps on the in

Página 115 - Alarm Trend Data

Bump Testing Multiple EAGLE 2s with Different Sensor Configurations • 196 Figure 138: Second GAS 2 FlowingThe PC Controller Program will analyze the

Página 116

197 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura- 16. After the last GAS 2 calibration gas has been applied, the docking station

Página 117 - Figure 82: Alarm Trend Data

Bump Testing Multiple EAGLE 2s with Different Sensor Configurations • 198 17. After the fresh air purge is complete, • If the bump test passes, the f

Página 118

199 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura-• If the bump test fails and the Force Calibration box was not selected, the fol

Página 119

Bump Testing Multiple EAGLE 2s with Different Sensor Configurations • 200• If the bump test fails and the Force Calibration box was selected, the SDM

Página 120

201 • Bump Testing Multiple EAGLE 2s with Different Sensor Configura-• If at any point during the bump test the flow rate to the instrument becomes t

Página 121

Calibrating Multiple EAGLE 2s with Different Sensor Configurations • 202Calibrating Multiple EAGLE 2s with Different Sensor ConfigurationsStandard EA

Página 122

203 • Calibrating Multiple EAGLE 2s with Different Sensor Configurations 5. Select the instruments you wish to calibrate and click on them to highlig

Página 123 - Calibration Data

Calibrating Multiple EAGLE 2s with Different Sensor Configurations • 204 7. You will be asked to confirm which EAGLE 2 channels will use GAS 1 for th

Página 124

Hardware Setup • 16Unit 2Unit 7Unit 4Unit 9Unit 1Unit 6Unit 5Unit 10Unit 3Unit 8 Figure 12: Gas 1 Tubing Connections

Página 125 - Event Data

205 • Calibrating Multiple EAGLE 2s with Different Sensor ConfigurationsNOTE: Do not select every gas in the gas cylinder confirmation screen unless

Página 126

Calibrating Multiple EAGLE 2s with Different Sensor Configurations • 206 8. You will then be asked to confirm the gas cylinder concentration and the

Página 127 - Interval Trend Data

207 • Calibrating Multiple EAGLE 2s with Different Sensor ConfigurationsIf you are calibrating single channel special sensor instruments and do not h

Página 128

Calibrating Multiple EAGLE 2s with Different Sensor Configurations • 208If you are calibrating single channel special sensor instruments and have a m

Página 129

209 • Calibrating Multiple EAGLE 2s with Different Sensor Configurations 9. Click OK. You will be asked if you want to calibrate the instruments. Cli

Página 130

Calibrating Multiple EAGLE 2s with Different Sensor Configurations • 210 11. After air is applied, and if all channels passed the zero adjustment, ca

Página 131

211 • Calibrating Multiple EAGLE 2s with Different Sensor Configurations 12. After the GAS 1 calibration gas is applied, calibration gas for the firs

Página 132

Calibrating Multiple EAGLE 2s with Different Sensor Configurations • 212 13. You will then be prompted to change the GAS 2 cylinder. The pumps on the

Página 133

213 • Calibrating Multiple EAGLE 2s with Different Sensor Configurations Figure 153: Second GAS 2 FlowingThe PC Controller Program will analyze the r

Página 134

Calibrating Multiple EAGLE 2s with Different Sensor Configurations • 214 15. After calibration gas is applied, the docking stations will purge with f

Página 135

17 • Hardware SetupNOTE: Use this figure if the EAGLE 2s that will be connected have exactly the same GAS 2 configuration.Unit 2Unit 7Unit 1Unit 6Uni

Página 136

215 • Calibrating Multiple EAGLE 2s with Different Sensor Configurations 16. After the fresh air purge is complete,• If the calibration passes, the f

Página 137 - Memo Data

Calibrating Multiple EAGLE 2s with Different Sensor Configurations • 216• If the calibration fails, the following window will open indicating which c

Página 138

217 • Calibrating Multiple EAGLE 2s with Different Sensor Configurations• If at any point during the calibration the flow rate to the instrument beco

Página 139 - Bump Test Data

Calibrating Multiple EAGLE 2s with Different Sensor Configurations • 218Appendix B: Bump Testing and Calibrating EAGLE 2s with IR SensorsThe default

Página 140

219 • Calibrating Multiple EAGLE 2s with Different Sensor ConfigurationsThe LEL only and autoranging ranges of the IR CH4 and IR “HC” sensors are the

Página 141

Hardware Setup • 18NOTE: Use this figure if 5 or less EAGLE 2s have exactly the same GAS 2 configuration.Unit 7Unit 4Unit 9Unit 1Unit 2Unit 6Unit 5Un

Página 142

19 • Hardware SetupConnecting Calibration Gas The GAS 1 and GAS 2 fittings on the back of the docking station are designed to be used with a calibrat

Página 143

Hardware Setup • 20(typically channel 6). If one of the two special sensors has a target gas that’s in a 5-gas cylinder (ie. SO2 or high range PID),

Página 144 - Instrument Function Menu

21 • EAGLE 2 and GX Type Instrument ConnectionTo connect calibration gas to the SDM-E2, do the following: 1. If the area around the docking station i

Página 145 - Edit Function

Installing the SDM-E2 Docking Station PC Controller Software • 22Installing the SDM-E2 Docking Station PC Controller Software 1. Launch Windows®. 2.

Página 146

23 • Installing the SDM-E2 Docking Station PC Controller Software 11. A Device Driver Window will appear prompting you to install necessary drivers.

Página 147

Installing the SDM-E2 Docking Station PC Controller Software • 24 14. Click Finish once the drivers are successfully installed. Figure 16: Finish De

Página 148

Table of ContentsTable of ContentsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Página 149

25 • Installing the SDM-E2 Docking Station PC Controller Software 18. The first time a SDM-E2 is turned on after being connected to the computer, a F

Página 150

Installing the SDM-E2 Docking Station PC Controller Software • 26 20. The wizard will search for the driver files. Figure 18: Searching 21. Once the

Página 151

27 • Launching the SDM-E2 Docking Station PC Controller ProgramLaunching the SDM-E2 Docking Station PC Controller Program 1. Click Start on the Windo

Página 152

Overview of the SDM-E2 Docking Station PC Controller Program • 28Overview of the SDM-E2 Docking Station PC Controller ProgramThis section provides a

Página 153

29 • Overview of the SDM-E2 Docking Station PC Controller ProgramMain Program WindowThe main program window is where the current status of the progra

Página 154 - Figure 112: Save As

Setting Up the PC Controller Program • 302s that are currently connected to the PC Controller Program and EAGLE 2s that are in the database. The SDM-

Página 155

31 • Setting Up the PC Controller Program 2. The configuration parameters are located in the Configuration Window of the SDM Docking Station PC Contr

Página 156

Setting Up the PC Controller Program • 32parameters in connected instruments. See “Edit Function” on page 140 for a complete description of the Edit

Página 157

33 • Setting Up the PC Controller Program 8. Select the Calibration options.In the Calibration section of the Parameter Tab, you can select Manual Ca

Página 158

Setting Up the PC Controller Program • 34If Auto Bump Test is selected, and an instrument is due for bump testing, a bump test will automatically beg

Página 159

Table of ContentsPC Controller Program Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28Main Program Window . . .

Página 160

35 • Setting Up the PC Controller Program• ToleranceThe Tolerance can be set between 10% and 50%. The factory setting is 50%. It determines how close

Página 161

Setting Up the PC Controller Program • 36 3. The Parameter Tab of the Configuration Window will be displayed. Figure 23: Config Window Parameter Tab

Página 162

37 • Setting Up the PC Controller Programorder on the Main Program Window, but the stations which are off will not be shown. 13. If you need to reord

Página 163

Setting Up the PC Controller Program • 38Selecting a Cylinder from the Predefined ListThe Cylinders Window comes with several predefined cylinders th

Página 164

39 • Setting Up the PC Controller ProgramCancel in the Cylinders Window. A window will appear notifying you that changes will not be saved. To contin

Página 165

Connecting Instruments to the PC Controller Program • 40 3. Type in the name of your new cylinder (ie. gas name and concentration). 4. Click in the P

Página 166

41 • Connecting Instruments to the PC Controller ProgramWARNING: Inserting the EAGLE 2 improperly may damage the exhaust bellow at the back of the SD

Página 167

Connecting Instruments to the PC Controller Program • 42LED will begin to flash amber. If the batteries are fully charged, the CHARGE LED will become

Página 168

43 • Connecting Instruments to the PC Controller Program 9. Turn on each EAGLE 2 by pressing and holding the POWER ENTER RESET button on each EAGLE 2

Página 169 - Figure 125: Channel Gas Tab

Connecting Instruments to the PC Controller Program • 44 Figure 29: EAGLE 2s Connected and DownloadedIf an instrument is due for calibration, the do

Página 170 - Clear Logs Function

Table of ContentsMemo Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132Bump Test Data.

Página 171 - Power Off Function

45 • Connecting Instruments to the PC Controller ProgramDetails View allows you to view connected EAGLE 2s and previously connected EAGLE 2s in a tab

Página 172

Connecting Instruments to the PC Controller Program • 46Printing and Exporting an Instrument List from the Main Program WindowYou can print, save, or

Página 173

47 • Instrument Placement for Bump Testing or CalibrationTo save the instrument list as a .csv file: 1. Click the Save icon located just below the Ex

Página 174 - C2H2Cl2F2

Instrument Placement for Bump Testing or Calibration • 48docking station manifold bank and the other set will be used for the second docking station

Página 175

49 • Automatic Bump Testing and CalibrationAutomatic Bump Testing and CalibrationThe PC Controller Program will automatically perform a bump test or

Página 176 - C4H2F5ClO

Bump Testing Standard 4-Gas Instrument(s) • 50Bump Testing Standard 4-Gas Instrument(s)The following instructions apply to EAGLE 2s with one or more

Página 177 - C6H18OSi2

51 • Bump Testing Standard 4-Gas Instrument(s) 5. Select the instrument or instruments you wish to bump test and click on them to highlight them as s

Página 178 - C3H2ClF5O

Bump Testing Standard 4-Gas Instrument(s) • 52 8. You will be asked to confirm which EAGLE 2 channels will use GAS 1 for the bump test. When using a

Página 179

53 • Bump Testing Standard 4-Gas Instrument(s) 9. You will then be asked to confirm the gas cylinder concentration and the instruments that the cylin

Página 180 - C2H2O2FNa

Bump Testing Standard 4-Gas Instrument(s) • 54 11. The pump will start and the program will indicate that air is flowing. The AIR indicator on the pr

Página 181 - C12H27O4P

1 • IntroductionIntroductionThis section briefly describes the SDM-E2 Docking Station, the Single Module Data Viewer Program, and the SDM Docking Sta

Página 182 - Config Window

55 • Bump Testing Standard 4-Gas Instrument(s) 12. After air is applied, and if all the channels passed the zero adjustment, calibration gas will be

Página 183

Bump Testing Standard 4-Gas Instrument(s) • 56 13. After calibration gas is applied, the docking station will purge with fresh air for 45 seconds. F

Página 184

57 • Bump Testing Standard 4-Gas Instrument(s) 14. After the fresh air purge is complete, • If the bump test passes, the following window will open i

Página 185

Bump Testing Standard 4-Gas Instrument(s) • 58• If the bump test fails and the Force Calibration box was not selected, the following window will open

Página 186 - Spare Parts List

59 • Bump Testing Standard 4-Gas Instrument(s)• If the bump test fails and the Force Calibration box was selected, the SDM-E2 automatically performs

Página 187

Bump Testing Standard 4-Gas Instrument(s) • 60• If at any point during the bump test the flow rate to the instrument becomes too low, the PC Controll

Página 188

61 • Calibrating a Standard 4-Gas InstrumentCalibrating a Standard 4-Gas InstrumentThe following instructions apply to EAGLE 2s with one or more of t

Página 189

Calibrating a Standard 4-Gas Instrument • 62 5. Select the instrument or instruments you wish to calibrate and click on them to highlight them as sho

Página 190

63 • Calibrating a Standard 4-Gas Instrument 7. You will be asked to confirm which EAGLE 2 channels will use GAS 1 for the calibration. When using a

Página 191

Calibrating a Standard 4-Gas Instrument • 64 8. You will then be asked to confirm the gas cylinder concentration and the instruments that the cylinde

Página 192

System Requirements • 2CAUTION: The EAGLE 2 detects oxygen deficiency and elevated levels of oxygen, combustible gases, carbon monoxide, and hydrogen

Página 193

65 • Calibrating a Standard 4-Gas Instrument 10. The pump will start and the program will indicate that air is flowing. The CAL S LED will begin to f

Página 194

Calibrating a Standard 4-Gas Instrument • 66 11. After air is applied, and if all the channels passed the zero adjustment, calibration gas will be ap

Página 195

67 • Calibrating a Standard 4-Gas Instrument 12. After calibration gas is applied, the docking station will purge with fresh air for 45 seconds. Fig

Página 196

Calibrating a Standard 4-Gas Instrument • 68 13. After the fresh air purge is complete,• If the calibration passes, the following window will open in

Página 197

69 • Calibrating a Standard 4-Gas Instrument• If the calibration fails, the following window will open indicating which channels failed to calibrate.

Página 198

Calibrating a Standard 4-Gas Instrument • 70• If at any point during the calibration the flow rate to the instrument becomes too low, the PC Controll

Página 199

71 • Bump Testing an Instrument with Special SensorsBump Testing an Instrument with Special SensorsStandard EAGLE 2s contain some combination of the

Página 200

Bump Testing an Instrument with Special Sensors • 72 4. Ensure that the gas out to EAGLE 2 line is connected to the inlet of the EAGLE 2. 5. Select t

Página 201

73 • Bump Testing an Instrument with Special Sensors 8. You will be asked to confirm which EAGLE 2 channels will use GAS 1 for the bump test. If you

Página 202

Bump Testing an Instrument with Special Sensors • 74 9. You will then be asked to confirm the gas cylinder concentration and the instruments that the

Página 203

3 • SpecificationsSpecificationsTable 1: SDM-E2 SpecificationsInput Power 12 VDCNOTE: AC Adapter with 100 - 240 VAC, 50/60 Hz, 0.6A input and 12 VDC,

Página 204

75 • Bump Testing an Instrument with Special SensorsIf you have a 5-gas cylinder that contains the special sensor’s target gas in the correct concent

Página 205

Bump Testing an Instrument with Special Sensors • 76If you are bump testing an instrument or instruments with 2 special sensors, the confirmation scr

Página 206

77 • Bump Testing an Instrument with Special Sensors 11. The pump will start and the program will indicate that air is flowing. The AIR indicator on

Página 207

Bump Testing an Instrument with Special Sensors • 78 12. After air is applied, and if all channels passed the zero adjustment, calibration gas from t

Página 208

79 • Bump Testing an Instrument with Special Sensors 13. After GAS 1 calibration gas is applied, calibration gas for the first special sensor will th

Página 209

Bump Testing an Instrument with Special Sensors • 80 14. If you are bump testing an instrument or instruments with 2 special sensors, you will then b

Página 210

81 • Bump Testing an Instrument with Special Sensors Figure 59: Second GAS 2 FlowingThe PC Controller Program will analyze the results. If the bump t

Página 211

Bump Testing an Instrument with Special Sensors • 82 15. After GAS 2 calibration gas has been applied, the docking station will purge with fresh air

Página 212

83 • Bump Testing an Instrument with Special Sensors 16. After the fresh air purge is complete, • If the bump test passes, the following window will

Página 213

Bump Testing an Instrument with Special Sensors • 84• If the bump test fails and the Force Calibration box was not selected, the following window wil

Página 214

About this Manual • 4About this ManualThe SDM-E2 Docking Station PC Controlled Configuration Operator’s Manual uses the following conventions for not

Página 215

85 • Bump Testing an Instrument with Special Sensors• If the bump test fails and the Force Calibration box was selected, the SDM-E2 automatically per

Página 216

Bump Testing an Instrument with Special Sensors • 86• If at any point during the bump test the flow rate to the instrument becomes too low, the PC Co

Página 217

87 • Calibrating an Instrument with Special SensorsCalibrating an Instrument with Special SensorsStandard EAGLE 2s contain some combination of the st

Página 218

Calibrating an Instrument with Special Sensors • 88 5. Select the instrument or instruments you wish to calibrate and click on them to highlight them

Página 219

89 • Calibrating an Instrument with Special Sensors 7. You will be asked to confirm which EAGLE 2 channels will use GAS 1 for the calibration. If you

Página 220

Calibrating an Instrument with Special Sensors • 90 8. You will then be asked to confirm the gas cylinder concentration and the instruments that the

Página 221

91 • Calibrating an Instrument with Special SensorsIf you have a 5-gas cylinder that contains the special sensor’s target gas in the correct concentr

Página 222

Calibrating an Instrument with Special Sensors • 92If you are calibrating an instrument or instruments with 2 special sensors, this confirmation scre

Página 223

93 • Calibrating an Instrument with Special Sensors 10. The pump will start and the program will indicate that air is flowing. The AIR indicator on t

Página 224

Calibrating an Instrument with Special Sensors • 94 11. After air is applied, and if all channels passed the zero adjustment, calibration gas from th