Rki-instruments GX-94 Manual do Utilizador Página 43

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 52
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 42
GX-94 Operator’s Manual Page 42
Lithium Memory Backup Battery
The memory backup battery is a 3.0 volt lithium battery which protects
against loss of the GX-94’s memory when the main batteries are dead
or removed. The memory backup battery typically lasts for 5 years.
Replace the memory backup battery every 5 years or if you notice that
the GX-94 is losing logged data when the main batteries are changed.
Figure 3: Lithium Backup Battery
1. Take the GX-94 to a non-hazardous location and turn the power
off.
2. Remove the four (4) phillips screws from the bottom of the instru-
ment. (Two in the belt clip and two next to the label.)
3. Pull the bottom and top case apart where the rubber cushion cov-
ers the seam. This opens the entire case.
4. Locate the lithium backup battery. Refer to Figure 3 above.
5. Gently pry the battery out of its holder with a small screw driver.
Memory Backup
Lithium Battery
GX-94 Shown w/out
Top Case
Vista de página 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 51 52

Comentários a estes Manuais

Sem comentários