Rki-instruments SDM-E2 Manual do Utilizador Página 59

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 58
Bump Testing Standard 4-Gas Instrument(s) • 54
11. The pump will start and the program will indicate that air is flowing. The AIR
indicator on the program screen will be on and air will be applied to the
instrument(s) for 45 seconds, then the PC Controller Program will perform a
zero adjustment on all the channels. The BUMP T LED on the SDM-E2
control panel will be flashing amber.
Figure 35: Air Flowing
The PC Controller Program will analyze the zero adjustment results. If any
sensors failed the zero adjustment, the bump test will be aborted.
Vista de página 58
1 2 ... 54 55 56 57 58 59 60 61 62 63 64 ... 223 224

Comentários a estes Manuais

Sem comentários