99 • Calibrating an Instrument with Special Sensors
15. After the fresh air purge is complete,
If the calibration passes, the following window will open indicating that. The
CAL S LED on the SDM-E2 control panel will be solid green. Click OK to
return to the main program window. The CAL S LED will turn off.
Figure 75: Calibration Passed
The EAGLE 2 will automatically be shut off 20 seconds after a successful
calibration.