Rki-instruments SDM-2009 PC Controlled Configuration Manual do Utilizador Página 51

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 50
Bump Testing a GX-2009 46
8. You will then be asked to confirm the auto calibration values and the
instruments that the calibration cylinder will be used for. Be sure the auto
calibration values match the values listed on the calibration cylinder. The serial
number and gas combination of all selected instruments will appear
sequentially beneath the gas cylinder concentration. All channels will be bump
tested using GAS 1.
Figure 32: Gas Order Confirmation Screen
9. Click OK. You will be asked if you want to bump test the instrument(s). Click
OK to proceed. If you don’t want to proceed with the bump test, click Cancel to
return to the main program window.
Vista de página 50
1 2 ... 46 47 48 49 50 51 52 53 54 55 56 ... 113 114

Comentários a estes Manuais

Sem comentários